Aoip Instrumentation TC6621 Thermocouple Calibrator

Technical Reference
Legislation and Compliance
RoHS Status: Not Applicable
COO (Country of Origin): FR
Product Details

Aoip Thermocouple Calibrator

The Aoip thermocouple calibrator with RS calibration is a handheld temperature device for thermocouples or resistive probes, equipped with a large graphical display and memory. It measures voltage ranging between -10mV and 100mV. This temperature calibrator measures or generates thermocouple temperatures (16 different types) and voltage with an accuracy of 0.02%. TC calibrators use a graphical interface, making programming and reading easier under graphical or test format. It is delivered with 4 alkaline AA batteries in the standard. The optional battery charger used a rechargeable battery.

Features and Benefits

High accuracy: 0.02% reading with an adjustable resolution of 1μV (TC 6621)
Values displayed in °C, °F, mV and ohm
Low-temperature coefficient: 10ppm/°C for thermocouples and 7ppm/°C for resistance
Measurements with HOLD function
Simulation of ramps, pre-programmed steps and synthesizer values
Correction of calibrated sensors
10,000 values stored and displayed graphically
Display resolution is selectable up to 0.1°C

Specifications
Attribute Value
Resolution 0.05 (E) °C, 0.05 (J) °C, 0.05 (K) °C, 0.05 (L) °C, 0.05 (Mo) °C, 0.05 (N) °C, 0.05 (NiMo/NiCo) °C, 0.05 (Plantine) °C, 0.05 (T) °C, 0.05 (U) °C, 0.1 (B) °C, 0.1 (C) °C, 0.1 (R) °C, 0.1 (S) °C, 0.2 (T) °C
Best Accuracy ±0.02% + 3 μV
Thermocouple Type B, C, E, J, K, L, N, R, S, T, U
Temperature Scale Centigrade
B Type Thermocouple Maximum Temperature +1820°C
B Type Thermocouple Minimum Temperature +400°C
E Type Thermocouple Maximum Temperature +1000°C
E Type Thermocouple Minimum Temperature -250°C
J Type Thermocouple Maximum Temperature +1200°C
K Type Thermocouple Maximum Temperature +1372°C
K Type Thermocouple Minimum Temperature -250°C
L Type Thermocouple Maximum Temperature +900°C
L Type Thermocouple Minimum Temperature -200°C
R Type Thermocouple Maximum Temperature +1768°C
U Type Thermocouple Maximum Temperature +600°C
U Type Thermocouple Minimum Temperature -200°C
Width 85mm
B Type Thermocouple Temperature Range +400 → +1820 °C
T Type Thermocouple Temperature Range -250 → +400 °C
J Type Thermocouple Temperature Range -210 → +1200 °C
Dimensions 157 x 85 x 45mm
S Type Thermocouple Temperature Range -50 → +1768 °C
S Type Thermocouple Minimum Temperature -50°C
R Type Thermocouple Temperature Range -50 → +1768 °C
S Type Thermocouple Maximum Temperature +1768°C
K Type Thermocouple Temperature Range -250 → +1372 °C
R Type Thermocouple Minimum Temperature -50°C
Length 157mm
U Type Thermocouple Temperature Range -200 → +600 °C
E Type Thermocouple Temperature Range -250 → +1000 °C
L Type Thermocouple Temperature Range -200 → +900 °C
T Type Thermocouple Minimum Temperature -250°C
T Type Thermocouple Maximum Temperature +400°C
Weight 306g
Model Number p TC6621
Height 45mm
6 In stock for delivery next working day
available from Europe for delivery within 2 working day(s)
Price Each
kr 7 387,77
(exc. VAT)
kr 9 234,71
(inc. VAT)
Units
Per unit
1 +
kr 7 387,77
Calibration Options:
Related Products
RS Pro TC920 is a great value, portable ...
Description:
RS Pro TC920 is a great value, portable instrument for calibrating and measuring type K thermocouples. 3 digit 2000 count LCDAdjustable -199.9 to +1230 °C temperature simulation and measure, ±(1 → 0.75% +2 → 1°C) low to high temperature ranges ...
A high performance portable instrument combining source and ...
Description:
A high performance portable instrument combining source and measure functions for thermocouples, Pt100, mV and mA. Measures/simulates 8 thermocouple types J, K, T, R, S, B, N, E: Pt100, mV and mADisplays in selectable units in °C/°F, uV/m. Large backlit ...
These compact, robust, easy-to-handle calibrators have large LCD ...
Description:
These compact, robust, easy-to-handle calibrators have large LCD screens and are suitable for use in highly diverse on-site conditions. They can calibrate an instrument in its operating environment and without removing it from the process. The principal functions are simulation ...